It is also possible that your web browser is not configured or not able to display style sheets. In this case, although the visual presentation will be degraded, the site should continue to be functional. We recommend using the latest version of Microsoft or Mozilla web browser to help minimise these problems.

Wiley InterScience

< Previous Abstract  |  Next Abstract >

Save Article to My Profile      Download Citation      Request Permissions

Abstract |  References  |  Full Text: HTML, PDF (Size: 72K)  | Related Articles | Citation Tracking

BLOOD COMPONENTS
In vitro and in vivo evaluation of LEUKOSEP HRC-600-C leukoreduction filtration system for red cells
James P. AuBuchon, Jose A. Cancelas, Louise Herschel, Jill Roger, Neeta Rugg, Patricia G. Pratt, Amy Joines, Jennifer F. Gormas, Majid Zia, and Michael R. Spearman
From the Cell Labeling Laboratory, Department of Pathology, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire; the Hoxworth Blood Center, University of Cincinnati Medical Center, Cincinnati, Ohio; and Hemerus Medical, LLC, St. Paul, Minnesota.
Correspondence to  James P. AuBuchon, Cell Labeling Laboratory, Department of Pathology, Dartmouth-Hitchcock Medical Center, Lebanon, NH 03756; e-mail: james.p.aubuchon@hitchcock.org.
Copyright 2006 American Association of Blood Banks

ABSTRACT

BACKGROUND: Documentation of the benefits of leukoreduction has led to the increased use of this technique and the need for development of efficient and effective techniques for its accomplishment. This study investigated the in vitro properties and in vivo autologous radiolabeled recovery of leukoreduced red cells (RBCs) produced through a leukoreduction filtration system for RBCs (LEUKOSEP HRC-600-C, Hemerus Medical).

STUDY DESIGN AND METHODS: Normal subjects donated 36 units of RBCs that were leukoreduced on Days 0, 3, or 5 through a "hands-off" technique. Biochemical studies were performed before and after filtration and at the end of 42 days of storage. Units leukoreduced on Days 0 or 5 were held until Day 42 and used for autologous radiolabeled return to determine recovery with 51Cr single-label radiolabeling techniques.

RESULTS: Leukoreduction filtration was accomplished in 16.3 ± 2 minutes on Day 0 at room temperature or 27 to 30 minutes on Days 3 or 5 after refrigeration. Leukoreduction efficiency was 4.6 ± 0.6 log with a median residual white blood cell (WBC) content of fewer than 3.3 × 104 WBCs per unit. RBC recovery was 90 ± 2 percent. Hemolysis was 0.34 ± 0.16 percent at the end of 42 days of storage. The in vivo recovery of radiolabeled RBCs 24 hours after autologous return was 80.6 ± 4.5 percent for RBC units leukoreduced on Days 0 and 5 combined.

CONCLUSION: The LEUKOSEP HRC-600-C WBC reduction filtration system produced leukoreduced RBCs efficiently and effectively with acceptable poststorage biochemical measures and posttransfusion recovery after 42 days of storage.


Received for publication November 30, 2005; revision received January 13, 2006, and accepted January 16, 2006.

DIGITAL OBJECT IDENTIFIER (DOI)
10.1111/j.1537-2995.2006.00897.x About DOI

Related Articles

  • Find other articles like this in Wiley InterScience
  • Find articles in Wiley InterScience written by any of the authors

Wiley InterScience is a member of CrossRef.

Cross Ref Member


Latest News & Information
TRF Impact Factor

Latest News & Information
TRF GSABC Submit Feature

Sign Up Now

Sign Up Now

Be the first to know about new research in your field

Sign up for FREE e-alerts from Wiley-Blackwell journals!

Sign Up Now